Abstract
Introduction: The limited resolution of the imaging system causes partial volume effects (PVEs). These results in spreading of image counts to the neighboring pixels. This phenomenon is called spill-out effect. This study aimed at quantifying PVEs using ImageJ. Methods: Technetium-99m solution of concentration of 74 kBq/ml was filled into spheres A, B and C of diameters: 26 mm, 20 mm and 16 mm respectively. The spheres were imaged mounted inside a Jaszczak phantom filled with activity free water using a Siemens E-Cam dual head gamma camera. Images were quantified using ImageJ following a two-step method. Step 1: Drawing of region of interest 1 (ROI 1) closely on the boundary of the planar image to extract images counts before PVEs correction. Step 2: Drawing region of interest 2 (ROI 2) to extract true sphere image counts. ROI 2 extends from the boundary of ROI 1 by the FWHM of the imaging system. Results: The study revealed that PVEs are aggravated by decrease in sphere size. Underestimation of image counts on the 64 × 64 pixels matrix was found to be: 9.7%; 15% and 26% in the order of decreasing sphere size. However, an improvement in the spatial resolution decreased PVEs (128 × 128 pixels: 6.7%; 12.0% and 22.5%; 256 × 256 pixels: 6.5%; 9% and 19.3%; 256 × 256 pixels: 6.1%; 8.0% and 18.7% in the order of decreasing sphere size). Conclusion: ImageJ successfully quantified PVEs attributed to the spill-out effect in planar imaging.